ARJOHUNTLEIGH POLSKA Sp. z.o.o. Recall 88469
Description: Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.
ARJOHUNTLEIGH POLSKA Sp. z.o.o. Recall 88469 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2588-2021 |
Event ID | 88469 |
Event Description | Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds. |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD. |
Quantity | 367 units |
Recall Reason | The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks. |
Device Classification | 20211006 |
Device Code Info | Manufacturing dates 5/25/2018-2/15/2021 (printed on both the bed and IndiGo module if the module was assembled on the bed at the time of manufacture, or printed on the module if it was later installed on the bed. |
Center Classification Date | 20210929 |
Recall Initiation Date | 20210610 |
Recalling Firm | ARJOHUNTLEIGH POLSKA Sp. z.o.o. |
Initial Notification | Letter |
Similar To |