Hill-Rom, Inc. Recall 88548
Description: Stretch Leveller Model Number: 3156200 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
Hill-Rom, Inc. Recall 88548 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2599-2021 |
Event ID | 88548 |
Event Description | Stretch Leveller Model Number: 3156200 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient. |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Latvia, Lithuania, Mexico, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates. |
Quantity | 122 units |
Recall Reason | Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from. |
Device Classification | 20211006 |
Device Code Info | N/A |
Center Classification Date | 20210930 |
Recall Initiation Date | 20210811 |
Recalling Firm | Hill-Rom, Inc. |
Initial Notification | Letter |