Philips North America Llc Recall 92992
Description: Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154
Philips North America Llc Recall 92992 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2608-2023 |
Event ID | 92992 |
Event Description | Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154 |
Product Type | Devices |
Distribution | Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Utd.Arab.Emir. Uzbekistan |
Quantity | 960 units |
Recall Reason | Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients |
Device Classification | 20230927 |
Device Code Info | UDI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 Software Version Number 4.x |
Center Classification Date | 20230919 |
Recall Initiation Date | 20230817 |
Recalling Firm | Philips North America Llc |
Initial Notification | Letter |
Similar To |