Product Safety Recalls

Product Recall Tracker

3M Company – Health Care Business Recall 92915

Description: 3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles

3M Company – Health Care Business Recall 92915 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2637-2023
Event ID92915
Event Description3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
Quantity79500 devices
Recall ReasonSpecific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
Device Classification20231004
Device Code InfoUDI/DI: 50707387762969 (shipper), 30707387762965 (carton); Lot codes: 33P3X3, exp. 12/29/2024; 33P4NH, exp. 01/01/2025
Center Classification Date20230926
Recall Initiation Date20230822
Recalling Firm3M Company - Health Care Business