Olympus Corporation of the Americas Recall 92893

Description: Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable

Olympus Corporation of the Americas Recall 92893 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2664-2023
Event ID92893
Event DescriptionVeran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Quantity60102 units
Recall ReasonNon-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Device Classification20231011
Device Code InfoINS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX 00815686020613; INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx 00815686020590; INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx 00815686020576; INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx 00815686020538; INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" 00815686020538; INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC 00815686020477; INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm 00815686020446; INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm 00815686020439; INS-0050 vPad Patient Tracker, 5/Bx 00815686020408; INS-0048 vPad Connector Cable 00815686020392; All serial numbers and all lot numbers
Center Classification Date20230929
Recall Initiation Date20230810
Recalling FirmOlympus Corporation of the Americas
Initial Notification N/A