Ion Beam Applications S.A. Recall 95080

Description: IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions

Ion Beam Applications S.A. Recall 95080 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-3089-2024
Event ID95080
Event DescriptionIBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions
Product TypeDevices
DistributionWorldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL, KA, IL, LA, MI, TN , TX, VA and the countries of BE, CZ, DE, ES, FR, IN, IT, JP, NL, PL, RU, SE, SG, TW.
Quantity35 units
Recall ReasonSafety Parameters Verification Mechanisms can be deactivated in clinical runtime
Device Classification20240918
Device Code InfoUDI-DI: 05404013801138 Serial Numbers: PAT.107 (DE), PAT.108 (US), PAT.109 (US), PAT.110 (US), PAT.111 (FR), PAT.112 (US), PAT.113 (US), PAT.114 (CZ), PAT.115 (IT), SAT.116 (US), SAT.117 (PL), SAT.118 (RU), SAT.119 (US), SAT.120 (DE), SAT.122 (SE), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (NL), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (FR), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.112 (BE), SBF.113 (US), SBF.117 (ES), SBF.124 (IT), SBF.125 (SG), SBF.128 (US), SBF.135 (US).
Center Classification Date20240910
Recall Initiation Date20240708
Recalling FirmIon Beam Applications S.A.
Initial Notification Letter