B Braun Medical Inc Recall 95171

Description: SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095

B Braun Medical Inc Recall 95171 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-3173-2024
Event ID95171
Event DescriptionSL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity10658351 units
Recall ReasonIncomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Device Classification20240925
Device Code InfoUDI-DI 04046964367762; All Unexpired Lot Numbers
Center Classification Date20240919
Recall Initiation Date20240815
Recalling FirmB Braun Medical Inc