Product Safety Recalls

Product Recall Tracker

Ondal AC2000 (PDI) Television Mounting System

Description: The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system. [Learn More]

215 distributed to 5 hospitals units were affected by this recall.

Ondal AC2000 (PDI) Television Mounting System
TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard

TRUMPF Medical Systems television mounting system.

Ondal AC2000 (PDI) Television Mounting System Recall Information

Country of OriginGermany
Recall Date07/18/2012
Recall ID1200
Recall Number12742
Product TypeTelevisions, Mounts, Guards & Antennas
HazardThe arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
InjuriesThe firm received one report of a mounting system falling while being positioned. No injuries have been reported.
RemedySecure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Remedy OptionRepair
ContactFor additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email [email protected]
Last Updated04/21/2016
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.